БПО-644 — resistance thermometer signal converter
Country: Ukraine · Model: БПО-644
error ±0.25 % · −200…+850 °C · output 0–5, 0–20, 4–20 mA / 0–10 V · supply 18–30 V DC
The БПО-644 is a single-channel analogue signal conditioner for resistance thermometers. It accepts copper (ТСМ 50М, 100М, гр.23), platinum (ТСП 50П, 100П, гр.21) and Pt sensors (Pt100, Pt500, Pt1000). The sensor is connected using a three-wire scheme. A resistance change within 2.5–3000 Ω is linearly converted into a standardised current signal of 0–5 mA, 0–20 mA or 4–20 mA, or into a 0–10 V voltage signal. The measuring range is −200…+850 °C and the error is ±0.25 %. The output is active. The unit runs on an 18–30 V DC supply and draws 30 mA. The housing mounts on a DIN rail and is rated IP20. It operates at ambient temperatures from −40 to +70 °C. Typical applications include power generation, metallurgy, the chemical industry, measurement systems and process automation. In these systems it feeds RTD signals to controllers, recorders and indicators.
- Number of channels
- 1
- Connection scheme
- three-wire
- Sensor types
- ТСМ (50М, 100М, гр.23), ТСП (50П, 100П, гр.21), Pt (100, 500, 1000)
- Input resistance range
- 2.5–3000 Ω
Technical specifications
| Number of channels | 1 |
|---|---|
| Connection scheme | three-wire |
| Sensor types | ТСМ (50М, 100М, гр.23), ТСП (50П, 100П, гр.21), Pt (100, 500, 1000) |
| Input resistance range | 2.5–3000 Ω |
| Measuring range | −200…+850 °C |
| Output signal | 0–5 mA (Rload ≤ 2000 Ω); 0–20 mA (Rload ≤ 500 Ω); 4–20 mA (Rload ≤ 500 Ω); 0–10 V (Rload ≥ 20 kΩ) |
| Output type | active |
| Error | ±0.25 % |
| Supply voltage | 18–30 V DC |
| Current consumption | 30 mA |
| Degree of protection | IP20 |
| Ambient temperature | −40…+70 °C |
| Mounting | DIN rail |
| Dimensions | 94 x 7 x 64 mm |
| Weight | 0.1 kg |
| Standards | ДСТУ EN 60529, ДСТУ IEC 60654-1:2001, ДСТУ IEC 60654-3:2001, ДСТУ 2887-94, ДСТУ 3272:2011, ДСТУ 2888-94 |
Specifications are taken from the manufacturer’s technical documentation. Updated .
Description
How to choose
Choose the БПО-644 if you need an active output (0–5, 0(4)–20 mA or 0–10 V) or if you use Pt1000 sensors. For a passive current loop, the related БПО-31 and БПО-430 models are available. They support Pt100/Pt500 and come in a larger 76x26x115 mm housing.
Similar products
Request a price and lead time
БПО-644 — resistance thermometer signal converter. We reply within one business day.
Or email office@usk.ua or call +38 (044) 339-50-20.
REQUEST
Need a price, lead time or replacement?
Send us a request and we will confirm the configuration, lead time and delivery terms. We reply by email.