Ukrainian Components WarehouseA USK project
UKRSKINDUSTRIAL COMPONENTS

TEST AND MEASUREMENT

Conductivity Sensor SC4AJ

The compact two-electrode SC4AJ conductivity sensor with a cell constant of 0.02 cm-1 or 0.1 cm-1 works at 0–110 °C and 0–1 MPa. It goes into boiler water loops, hydroponics and pure water lines.

Object of measurement
Conductivity of liquids or solutions
Measuring principle
Two-electrode system
Cell constant
0.02 cm-1, 0.1cm-1
Temperature Range
0 to 110℃ (For electrode)

Technical specifications

Object of measurementConductivity of liquids or solutions
Measuring principleTwo-electrode system
Cell constant0.02 cm-1, 0.1cm-1
Temperature Range0 to 110℃ (For electrode)
Sterilization for electrode135℃ (275℉), within 30 minutes in Steam Sterilization
Pressure range0 to 1 MPa (For electrode)
Sample solution conditionAlthough flow rate is not limited in measurement, air bubbles should not be mixed in the sample solutions to obtain correct measured values.
Temperature sensorPt1000
Process connectionAdapter mounting, welding socket, welding clamp

Specifications are taken from the manufacturer’s technical documentation. Updated .

Description

The sensor has two electrodes and a cell constant of 0.02 cm-1 or 0.1 cm-1. The small constant suits very pure water of low conductivity, the larger one middle values. The electrode withstands 0–110 °C and 0–1 MPa, and steam sterilization at 135 °C (275 °F) lasts up to 30 minutes, which matters for food and pharmaceutical lines. A Pt1000 is built in for temperature compensation. Sample flow is not limited, but air bubbles in it are not acceptable since they distort the result. It connects to the process through an adapter, a welding socket or a welding clamp. Besides boiler water and hydroponics, the sensor measures pure water resistivity in the semiconductor, food, pharmaceutical and power industries.

How to choose

The main selection parameter is the cell constant. Water of very low conductivity takes 0.02 cm-1 and higher values 0.1 cm-1. Next consider the connection, whether an adapter, welding socket or clamp is needed, and the temperature, since sterilization reaches 135 °C. If bubbles are possible in the stream, the mounting spot is chosen so that they do not enter the flow section.

Request a price and lead time

Conductivity Sensor SC4AJ. We reply within one business day.

* — required fields

Or email office@usk.ua or call +38 (044) 339-50-20.

FAQ

What cell constants does the SC4AJ have?

0.02 cm-1 and 0.1 cm-1.

What temperature and pressure does the SC4AJ electrode handle?

Temperature 0–110 °C and pressure 0–1 MPa. Steam sterilization at 135 °C is accepted for up to 30 minutes.

Which temperature sensor is built into the SC4AJ?

A Pt1000.

How is the SC4AJ connected to the process?

By adapter mounting, a welding socket or a welding clamp.

Do bubbles in the sample affect the SC4AJ?

Yes, air bubbles distort the readings, so there should be none. Sample flow itself is not limited.

Similar products

Test and measurement

REQUEST

Need a price, lead time or replacement?

Send us a request and we will confirm the configuration, lead time and delivery terms. We reply by email.

How we use your data

UKRSK, Ukrainian Components Warehouse, is a USK project. This form saves your email, request details, optional name and company, and attached files on the UKRSK server to process your enquiry. Files are not published in the catalogue. The form currently does not perform automated technical matching or send commercial quotations. To enquire about or delete your data, submit a request using this form and include your reference number. Do not attach passwords, payment details or unrelated confidential materials. Technical rate-limit records are stored separately; no advertising trackers are used.

To monitor page performance we collect LCP, INP and CLS measurements, language, screen category and a random measurement ID. These are stored on our hosting for up to 35 days; form contents, email, IP address and full URL are not recorded in this log. No cookies are used; Do Not Track and Global Privacy Control disable measurement.