Д5096, Д5097, Д5098 — laboratory AC/DC milliammeters

accuracy class 0.1 · 5–200 mA · 45–1500 Hz · weight max 4.3 kg

The Д5096, Д5097 and Д5098 are laboratory milliammeters for precise measurement of current in both AC and DC circuits. With an accuracy class of 0.1, they serve as reference instruments for the verification of less accurate ammeters and milliammeters. The series comprises three models that together cover full-scale range values from 5 to 200 mA. Each model offers two measuring limits. On alternating current the instruments operate across a frequency band of 45 to 1500 Hz. A normal and an extended working frequency range are specified for each model. Overall dimensions are (205 ± 1.45) x (290 ± 1.6) x (135 ± 2.0) mm, and the weight does not exceed 4.3 kg. Typical users are measurement and calibration laboratories that require high-accuracy monitoring of small currents and routine checking of working instruments.

Current type
AC and DC
Accuracy class
0.1
Full-scale range values
5–200 mA
Frequency range
45–1500 Hz

Technical specifications

Current typeAC and DC
Accuracy class0.1
Full-scale range values5–200 mA
Frequency range45–1500 Hz
Д5096: measuring limits5 and 10 mA
Д5096: frequency range (normal / working)45–300 Hz / over 300 to 1000 Hz
Д5097: measuring limits25 and 50 mA
Д5097: frequency range (normal / working)45–500 Hz / over 500 to 1500 Hz
Д5098: measuring limits100 and 200 mA
Д5098: frequency range (normal / working)45–500 Hz / over 500 to 1500 Hz
Overall dimensions(205 ± 1.45) x (290 ± 1.6) x (135 ± 2.0) mm
Weightmax 4.3 kg

Specifications are taken from the manufacturer’s technical documentation. Updated .

Description

How to choose

Select the model by the required current limit. The Д5096 covers 5 and 10 mA, with a normal frequency range of 45–300 Hz and a working range up to 1000 Hz. The Д5097 covers 25 and 50 mA, and the Д5098 covers 100 and 200 mA. Both have a normal range of 45–500 Hz and a working range up to 1500 Hz.

Similar products

Test and measurement

Request a price and lead time

Д5096, Д5097, Д5098 — laboratory AC/DC milliammeters. We reply within one business day.

* — required fields

Or email office@usk.ua or call +38 (044) 339-50-20.

REQUEST

Need a price, lead time or replacement?

Send us a request and we will confirm the configuration, lead time and delivery terms. We reply by email.

How we use your data

UKRSK, Ukrainian Components Warehouse, is a USK project. This form saves your email, request details, optional name and company, and attached files on the UKRSK server to process your enquiry. Files are not published in the catalogue. The form currently does not perform automated technical matching or send commercial quotations. To enquire about or delete your data, submit a request using this form and include your reference number. Do not attach passwords, payment details or unrelated confidential materials. Technical rate-limit records are stored separately; no advertising trackers are used.

To monitor page performance we collect LCP, INP and CLS measurements, language, screen category and a random measurement ID. These are stored on our hosting for up to 35 days; form contents, email, IP address and full URL are not recorded in this log. No cookies are used; Do Not Track and Global Privacy Control disable measurement.