Ukrainian Components WarehouseA USK project
UKRSKINDUSTRIAL COMPONENTS

TEST AND MEASUREMENT

Voltage and current calibrators МВ4303, МВ4305, МА4302, МА4304 (MV4303, MA4302)

Calibrators МВ4303, МВ4305, МА4302 and МА4304, often written as MV4303 or MA4302, reproduce DC voltage (МВ) and DC current (МА). They verify and calibrate working instruments of accuracy class 0,5 and higher, and serve as working standards.

Models
МВ4303, МВ4305 (voltage), МА4302, МА4304 (current)
Range МВ4303
1 mV – 1500 V DC
Subranges МВ4303
7
Resolution
from 1 µV

Technical specifications

ModelsМВ4303, МВ4305 (voltage), МА4302, МА4304 (current)
Range МВ43031 mV – 1500 V DC
Subranges МВ43037
Resolutionfrom 1 µV
Accuracy class0,1/0,002 … 0,2/0,06 by subrange
Load current МВ4303up to 100 mA
Settling timenot more than 2 s
Warm-up30 min
Continuous operationup to 16 h
Power consumptionup to 250 V·A
Dimensions МВ4303485 × 510 × 215 mm
Other spellingsMV4303, MV4305, MA4302, MA4304

Specifications are taken from the manufacturer’s technical documentation. Updated .

Description

МВ4303 reproduces DC voltage from 1 mV to 1500 V in seven subranges, with resolution from 1 µV. Accuracy class is 0,1/0,002 to 0,2/0,06 depending on subrange. Load current reaches 100 mA. The instrument shows the error of the device under test and settles within 2 s after setting. Warm-up takes 30 min; continuous operation lasts up to 16 h. Power consumption is up to 250 V·A; dimensions 485 × 510 × 215 mm. Used by metrology services and verification labs to calibrate voltmeters, ammeters and multimeters, and by instrument manufacturers.

How to choose

Start with the quantity to reproduce: voltage needs МВ4303 or МВ4305, current needs МА4302 or МА4304. Then match the range and accuracy class to the instruments under test. Confirm kit, verification and version with a USK ENERGY manager when ordering.

Request a price and lead time

Voltage and current calibrators МВ4303, МВ4305, МА4302, МА4304 (MV4303, MA4302). We reply within one business day.

* — required fields

Or email office@usk.ua or call +38 (044) 339-50-20.

FAQ

Are МВ4303 and MV4303 one calibrator?

Yes. Cyrillic МВ and Latin MV denote the same instrument, as МА4302 and MA4302 do.

What is the difference between МВ and МА?

МВ calibrators reproduce DC voltage, МА calibrators DC current.

What is the range of МВ4303?

1 mV to 1500 V DC in seven subranges, resolution from 1 µV.

Which instruments do these calibrators suit?

Verification and calibration of instruments of accuracy class 0,5 and higher, such as voltmeters, ammeters and multimeters.

Is a warm-up needed?

Yes, 30 min for МВ4303. Afterwards it can run continuously for up to 16 h.

Similar products

Test and measurement

REQUEST

Need a price, lead time or replacement?

Send us a request and we will confirm the configuration, lead time and delivery terms. We reply by email.

How we use your data

UKRSK, Ukrainian Components Warehouse, is a USK project. This form saves your email, request details, optional name and company, and attached files on the UKRSK server to process your enquiry. Files are not published in the catalogue. The form currently does not perform automated technical matching or send commercial quotations. To enquire about or delete your data, submit a request using this form and include your reference number. Do not attach passwords, payment details or unrelated confidential materials. Technical rate-limit records are stored separately; no advertising trackers are used.

To monitor page performance we collect LCP, INP and CLS measurements, language, screen category and a random measurement ID. These are stored on our hosting for up to 35 days; form contents, email, IP address and full URL are not recorded in this log. No cookies are used; Do Not Track and Global Privacy Control disable measurement.