Ф5043 — laboratory AC frequency meter
Country: Ukraine · Model: Ф5043
20–20000 Hz · accuracy classes 0.1; 0.2; 0.5 · input voltage 1–500 V · consumption up to 13 V·A
The Ф5043 is a laboratory frequency meter for measuring AC frequency from 20 to 20000 Hz. It accepts sinusoidal signals and pulse signals with a duration of at least 25 μs. It also handles signals of arbitrary shape, provided they have no more than two extrema per period. The nominal input voltage range is 1 to 500 V. Measurement is divided into 21 narrow and 7 wide ranges, so the operator can select the most suitable span for each task. Depending on the range selected, the intrinsic error limits are ±0.1 %, ±0.2 % or ±0.5 % of the range end value. These correspond to accuracy classes 0.1, 0.2 and 0.5. Power drawn from the mains does not exceed 13 V·A. Overall dimensions are no more than 160x210x370 mm, and the instrument weighs up to 8 kg. The meter is intended for laboratory measurements, such as checking the frequency of AC sources and verifying other measuring equipment.
- Measuring frequency range
- 20–20000 Hz
- Narrow measuring ranges (21)
- 25-35, 35-45, 45-55, 50-70, 70-90, 90-110, 125-175, 175-225, 225-275, 250-350, 350-450, 450-550, 500-700, 700-900, 900-1100, 1250-1750, 1750-2250, 2250-2750, 2500-3500, 3500-4500, 4500-5500 Hz
- Wide measuring ranges (7)
- 20-200, 40-400, 100-1000, 200-2000, 400-4000, 1000-10000, 2000-20000 Hz
- Accuracy classes
- 0.1; 0.2; 0.5
Technical specifications
| Measuring frequency range | 20–20000 Hz |
|---|---|
| Narrow measuring ranges (21) | 25-35, 35-45, 45-55, 50-70, 70-90, 90-110, 125-175, 175-225, 225-275, 250-350, 350-450, 450-550, 500-700, 700-900, 900-1100, 1250-1750, 1750-2250, 2250-2750, 2500-3500, 3500-4500, 4500-5500 Hz |
| Wide measuring ranges (7) | 20-200, 40-400, 100-1000, 200-2000, 400-4000, 1000-10000, 2000-20000 Hz |
| Accuracy classes | 0.1; 0.2; 0.5 |
| Intrinsic error, % of range end value | ±0.1 (45-55 Hz and multiples); ±0.2 (25-35, 35-45 Hz and multiples); ±0.5 (wide ranges) |
| Nominal input voltage range | 1–500 V |
| Input signals | sinusoidal; pulse, duration at least 25 μs; arbitrary shape with no more than two extrema per period |
| Power consumption | max. 13 V·A |
| Overall dimensions | max. 160x210x370 mm |
| Weight | max. 8 kg |
| Scope of supply | frequency meter, cable, tip (2 pcs), clamp (2 pcs), fuse ПМ 0,25 (2 pcs), indicator lamp ИНС-1, technical description and operating manual, passport |
Specifications are taken from the manufacturer’s technical documentation. Updated .
Description
How to choose
For the highest accuracy (±0.1 %), use the narrow 45–55 Hz range or one of its multiples. The 25–35 Hz and 35–45 Hz ranges and their multiples give ±0.2 %. If the frequency is unknown in advance or varies widely, use one of the seven wide ranges (from 20–200 to 2000–20000 Hz), which have an error of ±0.5 %.
Similar products
Request a price and lead time
Ф5043 — laboratory AC frequency meter. We reply within one business day.