И561 — laboratory current transformer, accuracy class 0.02

accuracy class 0.02 · 0.5–100 A / 5 A · 45–65 Hz · 660 V

The И561 is a laboratory current transformer for single-phase AC circuits at 50 Hz. It converts a primary current of 0.5 to 100 A into a standardised secondary current of 5 A. This allows standard measuring instruments to be connected and high-precision comparisons to be made. Permissible error limits correspond to accuracy class 0.02. The unit therefore serves as a reference standard for verifying current transformers and for precise electrical measurements in laboratories and metrology departments. The primary winding offers 13 rated current values, and the measuring range is 2–120% of the rated value. The rated voltage of the primary winding relative to the case and secondary winding is 660 V. The rated secondary load is 5 VA at a power factor of 1, and the normal frequency range is 45–65 Hz. Continuous operation at rated current is limited to 8 h. Overall dimensions do not exceed 400 x 300 x 130 mm, and the weight is no more than 25 kg.

Type
laboratory current transformer, single-phase
Accuracy class
0.02
Rated primary currents
0.5; 1; 2.5; 5; 10; 15; 20; 30; 40; 50; 60; 80; 100 A
Primary current measuring range
2–120% of rated value

Technical specifications

Typelaboratory current transformer, single-phase
Accuracy class0.02
Rated primary currents0.5; 1; 2.5; 5; 10; 15; 20; 30; 40; 50; 60; 80; 100 A
Primary current measuring range2–120% of rated value
Rated secondary current5 A
Rated voltage of primary winding to case and secondary winding660 V
Rated secondary load5 VA at power factor 1
Normal frequency range45–65 Hz
Power consumption at rated primary current 0.5 / 100 Amax. 15 / 40 VA
Continuous operation at rated currentmax. 8 h
Ratio of operating time to rest intervalmax. 4
Overall dimensionsmax. 400 x 300 x 130 mm
Weightmax. 25 kg

Specifications are taken from the manufacturer’s technical documentation. Updated .

Description

How to choose

Select the rated primary current from the series 0.5, 1, 2.5, 5, 10, 15, 20, 30, 40, 50, 60, 80 and 100 A. Base the choice on the expected current of the transformer under test or of the measuring circuit. That current should fall within 2–120% of the chosen rating. The secondary current remains 5 A for every rating.

Similar products

Test and measurement

Request a price and lead time

И561 — laboratory current transformer, accuracy class 0.02. We reply within one business day.

* — required fields

Or email office@usk.ua or call +38 (044) 339-50-20.

REQUEST

Need a price, lead time or replacement?

Send us a request and we will confirm the configuration, lead time and delivery terms. We reply by email.

How we use your data

UKRSK, Ukrainian Components Warehouse, is a USK project. This form saves your email, request details, optional name and company, and attached files on the UKRSK server to process your enquiry. Files are not published in the catalogue. The form currently does not perform automated technical matching or send commercial quotations. To enquire about or delete your data, submit a request using this form and include your reference number. Do not attach passwords, payment details or unrelated confidential materials. Technical rate-limit records are stored separately; no advertising trackers are used.

To monitor page performance we collect LCP, INP and CLS measurements, language, screen category and a random measurement ID. These are stored on our hosting for up to 35 days; form contents, email, IP address and full URL are not recorded in this log. No cookies are used; Do Not Track and Global Privacy Control disable measurement.