Ukrainian Components WarehouseA USK project
UKRSKINDUSTRIAL COMPONENTS

TEST AND MEASUREMENT

ІПМ-2Н — single-phase AC voltage indicator-transducer

accuracy ≤ 0.4 % · 1÷10 V … 50÷500 V · outputs 0..5 mA, 0..20 mA, 4..20 mA, 0..10 V · RS-485, Modbus RTU

The ІПМ-2Н is a microprocessor-based indicator-transducer for AC voltage in single-phase networks. It combines a digital voltmeter with a signal conditioning transducer. The unit measures RMS voltage and shows the value on a 4-digit display. At the same time it generates a standard analogue signal of 0..5 mA, 0..20 mA, 4..20 mA or 0..10 V. Depending on the version, discrete outputs for threshold alarms are available: an NPN transistor or a 230 V, 5 A relay. An optional RS-485 interface with Modbus RTU links the device to SCADA and HMI systems. Input ranges run from 1÷10 V to 50÷500 V, and connection via a transformer is also possible. Accuracy is ≤ 0.4 %, the measurement period is 0.1 s and galvanic isolation is at least 1500 V. The IP30 housing measures 98 × 54 × 57 mm, mounts on a DIN rail and operates from -40 °C to +70 °C. Typical applications are in the energy sector, metallurgy, and the chemical and food industries.

Measured quantity
single-phase AC voltage (RMS)
Input signal ranges
1÷10 V; 1.5÷15 V; 3÷30 V; 4÷40 V; 5÷50 V; 6÷60 V; 10÷100 V; 25÷250 V; 30÷300 V; 50÷500 V; from transformer
Accuracy
≤ 0.4 %
Measurement period
0.1 s

Technical specifications

Measured quantitysingle-phase AC voltage (RMS)
Input signal ranges1÷10 V; 1.5÷15 V; 3÷30 V; 4÷40 V; 5÷50 V; 6÷60 V; 10÷100 V; 25÷250 V; 30÷300 V; 50÷500 V; from transformer
Accuracy≤ 0.4 %
Measurement period0.1 s
Display4-digit display
Analogue output0..5 mA; 0..20 mA; 4..20 mA; 0..10 V
Discrete outputNPN transistor; relay 230 V, 5 A
InterfaceRS-485 (Modbus RTU)
Galvanic isolation≥ 1500 V
Supply voltage230 V; 24 V
Power consumption≤ 2.5 VA; ≤ 110 mA
MountingDIN rail (DIN35x7.5)
Protection degreeIP30
Operating temperature-40 °C … +70 °C
Dimensions98 mm x 54 mm x 57 mm
Weight< 0.15 kg
StandardsДСТУ IEC 60381-1; ДСТУ IEC 60654-1:2001; ДСТУ IEC 60654-3:2001; ДСТУ EN 60529:2014; ДСТУ EN 61326-1

Specifications are taken from the manufacturer’s technical documentation. Updated .

Description

How to choose

Select the input range closest to the voltage being monitored: from 1÷10 V to 50÷500 V. Higher levels can be connected via a transformer. Next, choose the analogue output type (0..5 mA, 0..20 mA, 4..20 mA, 0..10 V) and the discrete output type (NPN or 230 V, 5 A relay). Specify whether RS-485 is required and select a supply voltage of 230 V or 24 V. Example order code: ІПМ-2Н-03-3-P-1-230B — 3–30 V input, 4–20 mA output, relay, RS-485, 230 V supply.

Similar products

Test and measurement

Request a price and lead time

ІПМ-2Н — single-phase AC voltage indicator-transducer. We reply within one business day.

* — required fields

Or email office@usk.ua or call +38 (044) 339-50-20.

REQUEST

Need a price, lead time or replacement?

Send us a request and we will confirm the configuration, lead time and delivery terms. We reply by email.

How we use your data

UKRSK, Ukrainian Components Warehouse, is a USK project. This form saves your email, request details, optional name and company, and attached files on the UKRSK server to process your enquiry. Files are not published in the catalogue. The form currently does not perform automated technical matching or send commercial quotations. To enquire about or delete your data, submit a request using this form and include your reference number. Do not attach passwords, payment details or unrelated confidential materials. Technical rate-limit records are stored separately; no advertising trackers are used.

To monitor page performance we collect LCP, INP and CLS measurements, language, screen category and a random measurement ID. These are stored on our hosting for up to 35 days; form contents, email, IP address and full URL are not recorded in this log. No cookies are used; Do Not Track and Global Privacy Control disable measurement.