The М2005 is a portable multi-range microammeter of the magnetoelectric system. It is designed to measure small currents in DC circuits. The instrument has seven ranges with full-scale values of 10, 25, 50, 100, 250, 500 and 1000 мкА. The permissible basic error on every range is ±0.2%, which corresponds to accuracy class 0.2. The voltage drop across the instrument depends on the selected range and lies between 63 мВ and 624 мВ. Readings settle within 4 s at most. The insulation is tested at 2 кВ. The meter is rated for ambient temperatures from +10 to +35°C and relative humidity up to 80% at +25°C. Overall dimensions do not exceed 243 x 200 x 100 мм. It weighs 2.6 кг without its case and 3.7 кг with the case. Laboratories and industrial test departments use it for precise current measurement in DC circuits, including calibration and commissioning work.

System
magnetoelectric
Design
portable, multi-range
Current type
DC
Accuracy class
0.2

Technical specifications

Systemmagnetoelectric
Designportable, multi-range
Current typeDC
Accuracy class0.2
Permissible basic error±0.2% on all ranges
Full-scale values of ranges10, 25, 50, 100, 250, 500, 1000 мкА
Voltage drop, 10 мкА range63 мВ
Voltage drop, 25 мкА range400 мВ
Voltage drop, 50 мкА range516 мВ
Voltage drop, 100 мкА range573 мВ
Voltage drop, 250 мкА range607 мВ
Voltage drop, 500 мкА range618 мВ
Voltage drop, 1000 мкА range624 мВ
Settling timemax. 4 s
Insulation test voltage2 кВ
Ambient temperature+10 to +35°C
Relative humidityup to 80% at +25°C
Overall dimensionsmax. 243 x 200 x 100 мм
Weight without casemax. 2.6 кг
Weight with casemax. 3.7 кг
Scope of supplymicroammeter, case, technical description and operating manual, passport

Specifications are taken from the manufacturer’s technical documentation. Updated .

Description

How to choose

All seven ranges (10–1000 мкА) are built into one instrument. Select the lowest range that covers the expected current to make full use of the accuracy. Take the voltage drop into account: it is 63 мВ on the 10 мкА range and 624 мВ on the 1000 мкА range. This matters in low-voltage circuits.

Similar products

Test and measurement

Request a price and lead time

М2005 — portable multi-range DC microammeter. We reply within one business day.

* — required fields

Or email office@usk.ua or call +38 (044) 339-50-20.

REQUEST

Need a price, lead time or replacement?

Send us a request and we will confirm the configuration, lead time and delivery terms. We reply by email.

How we use your data

UKRSK, Ukrainian Components Warehouse, is a USK project. This form saves your email, request details, optional name and company, and attached files on the UKRSK server to process your enquiry. Files are not published in the catalogue. The form currently does not perform automated technical matching or send commercial quotations. To enquire about or delete your data, submit a request using this form and include your reference number. Do not attach passwords, payment details or unrelated confidential materials. Technical rate-limit records are stored separately; no advertising trackers are used.

To monitor page performance we collect LCP, INP and CLS measurements, language, screen category and a random measurement ID. These are stored on our hosting for up to 35 days; form contents, email, IP address and full URL are not recorded in this log. No cookies are used; Do Not Track and Global Privacy Control disable measurement.