М2044, М2051 — laboratory DC voltammeters
Country: Ukraine · Model: М2044, М2051
accuracy class 0.2 / 0.5 · 0.75 mA – 30 A · 15 mV – 600 V · insulation test voltage 2 kV
The М2044 and М2051 are laboratory voltammeters for measuring current and voltage in DC circuits. A single instrument replaces a set of separate ammeters and voltmeters. The required range is set with a range selector and a multiplier switch (Х1, Х2, Х4). Current ranges run from 0.75 mA to 3000 mA and from 7.5 A to 30 A. Voltage ranges run from 15 mV to 600 V. The limit of intrinsic error on all ranges is ±0.2 % for the М2044 and ±0.5 % for the М2051. Reading settling time is no more than 6 s on the 15 mV range and no more than 4 s on all other ranges. Insulation is tested at 2 kV, mean time between failures is 32500 h and mean service life is 10 years. Calibrated connecting leads and a carrying case are supplied. Typical applications are measurement laboratories, DC circuit testing, and the calibration and verification of measuring instruments.
- Current type
- DC
- Accuracy class
- 0.2 (М2044); 0.5 (М2051)
- Intrinsic error limit
- ±0.2 % (М2044); ±0.5 % (М2051)
- Current ranges, mA
- 0.75; 1.5; 3; 7.5; 15; 30; 75; 150; 300; 750; 1500; 3000
Technical specifications
| Current type | DC |
|---|---|
| Accuracy class | 0.2 (М2044); 0.5 (М2051) |
| Intrinsic error limit | ±0.2 % (М2044); ±0.5 % (М2051) |
| Current ranges, mA | 0.75; 1.5; 3; 7.5; 15; 30; 75; 150; 300; 750; 1500; 3000 |
| Current ranges, A | 7.5; 15; 30 |
| Voltage ranges, mV | 15; 30; 60; 75; 150; 300 |
| Voltage ranges, V | 0.75; 1.5; 3; 7.5; 15; 30; 75; 150; 300; 600 |
| Multiplier switch | Х1, Х2, Х4 |
| Voltage drop on current ranges, max | 20–300 mV (depending on range) |
| Full-scale current on voltage ranges | 0.25; 0.5; 1 mA |
| Reading settling time | max 6 s (15 mV); max 4 s (other ranges) |
| AC ripple factor | 1 % (М2044); 3 % (М2051) |
| Additional error per 10 °C | ±0.2 % (М2044); ±0.4 % (М2051) |
| Additional error from 5° tilt | ±0.2 % (М2044); ±0.5 % (М2051) |
| Additional error from external magnetic field | ±0.3 % |
| Additional error from ferromagnetic base | ±0.1 % |
| Additional error from adjacent instrument | ±0.15 % |
| Insulation test voltage | 2 kV |
| Mean time between failures | 32500 h |
| Mean time to repair | 6 h |
| Mean service life | 10 years |
| Operating temperature | 10 to 35 °C |
| Relative humidity | up to 80 % at 25 °C |
| Reference conditions: temperature | 20±2 °C |
| Reference conditions: humidity | 30–80 % |
| Atmospheric pressure | 84–106.7 kPa |
| Overall dimensions, max | 243×200×100 mm |
| Weight without case | 3 kg |
| Weight with case | 4.1 kg |
| Delivery set | calibrated connecting leads (0.035 ± 0.002) Ω — 1 pair; carrying case; operating manual |
Specifications are taken from the manufacturer’s technical documentation. Updated .
Description
How to choose
Both models share the same measuring ranges, dimensions and delivery set. The М2044 (class 0.2) is rated for signals with an AC ripple factor of up to 1 % and suits verification work and precise laboratory measurement. The М2051 (class 0.5) tolerates ripple of up to 3 % and suits routine measurement in DC circuits.