НПМ-612 — universal single-channel signal converter

accuracy ±0.2 % · outputs 0–5, 0–20, 4–20 mA, 0–10 V · supply 18–36 V DC · measurement period 0.1 s

The НПМ-612 is a microprocessor-based single-channel signal converter with a universal input. It accepts thermocouples (K, J, T, E, N, S, R, B, L), resistance thermometers (Pt50, Pt100, Pt500, Pt1000, TSP50, TSP100, TSM50, TSM100), unified current and voltage signals, and resistance up to 2500 Ohm. The unit linearises and scales the measured value and delivers a standard analogue output of 0–5 mA, 0–20 mA, 4–20 mA or 0–10 V. Input, supply and output are galvanically isolated, with isolation rated up to 500 V. Accuracy is ±0.2 % and the measurement period is 0.1 s. It runs on 24 V (18–36 V DC) and draws 60 mA. Configuration is carried out over USB with configuration software. Settings are stored in non-volatile memory, and Modbus RTU is supported. The 91x15x64 mm housing mounts on a DIN rail. Typical uses are temperature measurement and pressure and flow monitoring in industrial automation systems, including the food industry.

Input signals: current
0–5 mA, 0–20 mA, 4–20 mA, -5…5 mA, -20…20 mA
Input signals: voltage
0–10 V, 0–100 mV, -10…10 V, -100…100 mV
Resistance thermometers
Pt50, Pt100, Pt500, Pt1000, TSP50, TSP100, TSM50, TSM100
Thermocouples
K, J, T, E, N, S, R, B, L

Technical specifications

Input signals: current0–5 mA, 0–20 mA, 4–20 mA, -5…5 mA, -20…20 mA
Input signals: voltage0–10 V, 0–100 mV, -10…10 V, -100…100 mV
Resistance thermometersPt50, Pt100, Pt500, Pt1000, TSP50, TSP100, TSM50, TSM100
ThermocouplesK, J, T, E, N, S, R, B, L
Input signals: resistance0–150 Ohm, 0–300 Ohm, 0–1200 Ohm, 0–2500 Ohm
Output signals0–5 mA, 0–20 mA, 4–20 mA, 0–10 V
Number of channels1
Accuracy±0.2 %
Measurement period0.1 s
Supply voltage24 V (18–36 V DC)
Current consumption60 mA
Galvanic isolationup to 500 V
InterfaceUSB
ProtocolModbus RTU
MountingDIN rail (DIN35x7,5 EN50022)
Protection ratingIP20
Operating temperature-40…+70 °C
Dimensions91x15x64 mm
Weight0.3 kg
StandardsIEC 381-1, IEC 60381-2

Specifications are taken from the manufacturer’s technical documentation. Updated .

Description

How to choose

The version is defined by an input signal code (01–31) and an output signal code (01–04). For example, НПМ-612-07-2 has a 4–20 mA input and a 0–20 mA output. Choose the input code to suit the sensor type (thermocouple, RTD, current, voltage or resistance). Choose the output code to suit the receiving instrument or controller. Note that some versions have a non-isolated output.

Similar products

Test and measurement

Request a price and lead time

НПМ-612 — universal single-channel signal converter. We reply within one business day.

* — required fields

Or email office@usk.ua or call +38 (044) 339-50-20.

REQUEST

Need a price, lead time or replacement?

Send us a request and we will confirm the configuration, lead time and delivery terms. We reply by email.

How we use your data

UKRSK, Ukrainian Components Warehouse, is a USK project. This form saves your email, request details, optional name and company, and attached files on the UKRSK server to process your enquiry. Files are not published in the catalogue. The form currently does not perform automated technical matching or send commercial quotations. To enquire about or delete your data, submit a request using this form and include your reference number. Do not attach passwords, payment details or unrelated confidential materials. Technical rate-limit records are stored separately; no advertising trackers are used.

To monitor page performance we collect LCP, INP and CLS measurements, language, screen category and a random measurement ID. These are stored on our hosting for up to 35 days; form contents, email, IP address and full URL are not recorded in this log. No cookies are used; Do Not Track and Global Privacy Control disable measurement.