Ukrainian Components WarehouseA USK project
UKRSKINDUSTRIAL COMPONENTS

TEST AND MEASUREMENT

Р4830/1, Р4830/2, Р4830/3 — laboratory resistance boxes for DC and AC circuits

accuracy class 0,05/2,5·10-5 to 0,05/2,5·10-7 · 0,01–1222221 Ω · up to 50 kHz · inductance ≤ 1μH

The Р4830/1, Р4830/2 and Р4830/3 are laboratory decade resistance boxes. They act as a standard of electrical resistance in DC and AC circuits at frequencies up to 50 kHz. The required value is set with decade switches. With every switch at zero, a residual resistance remains: 0,03 Ω through the ordinary terminals and 0,02 Ω through the low-ohmic terminals (0,015 Ω on the Р4830/2). The three versions differ in range and resolution. The Р4830/1 covers 0,01–12222,21 Ω in 0,01 Ω steps. The Р4830/2 spans 0,1–122222,1 Ω in 0,1 Ω steps. The Р4830/3 covers 1,0–1222221 Ω in 1,0 Ω steps. Their accuracy classes are 0,05/2,5·10-5, 0,05/2,5·10-6 and 0,05/2,5·10-7 respectively. When resistance up to 1 Ω is switched in through the low-ohmic terminals, inductance does not exceed 1μH. The boxes are used in measurement laboratories for precision measurement, testing and calibration, and for setting up DC and AC measuring circuits.

Purpose
standard of electrical resistance in DC and AC circuits
Accuracy class Р4830/1
0,05/2,5·10-5
Accuracy class Р4830/2
0,05/2,5·10-6
Accuracy class Р4830/3
0,05/2,5·10-7

Technical specifications

Purposestandard of electrical resistance in DC and AC circuits
Accuracy class Р4830/10,05/2,5·10-5
Accuracy class Р4830/20,05/2,5·10-6
Accuracy class Р4830/30,05/2,5·10-7
Resistance range Р4830/10,01 to 12222,21 Ω, step 0,01 Ω
Resistance range Р4830/20,1 to 122222,1 Ω, step 0,1 Ω
Resistance range Р4830/31,0 to 1222221 Ω, step 1,0 Ω
Initial resistance via ordinary terminals «1»—«3»0,03 Ω
Initial resistance via low-ohmic terminals «2»—«3»0,02 Ω (Р4830/1, Р4830/3); 0,015 Ω (Р4830/2)
Initial resistance variation0,003 Ω (ordinary terminals); 0,002 Ω, 0,0015 Ω for Р4830/2 (low-ohmic terminals)
Inductance (up to 1 Ω, via low-ohmic terminals)≤ 1μH
Maximum frequency50 kHz
Upper frequency limit per decade1 Ω — 10,0 kHz; 10 and 10² Ω — 50,0 kHz; 10³ Ω — 10,0 kHz; 10⁴ Ω — 5,0 kHz; 10⁵ Ω — 0,5 kHz
Insulation resistance Р4830/1≥ 2·10^8 Ω
Insulation resistance Р4830/2≥ 2·10^9 Ω
Insulation resistance Р4830/3≥ 2·10^10 Ω
Overall dimensionsmax. 490х135х365 mm
Weightmax. 8,5 kg

Specifications are taken from the manufacturer’s technical documentation. Updated .

Description

How to choose

Choose the version by the range and resolution you need. The Р4830/1 suits low resistances, up to 12222,21 Ω in 0,01 Ω steps. The Р4830/2 covers mid-range values, up to 122222,1 Ω in 0,1 Ω steps. The Р4830/3 is for high-resistance work, up to 1222221 Ω in 1,0 Ω steps. For AC work, check the upper frequency limit of the decade in use.

Similar products

Test and measurement

Request a price and lead time

Р4830/1, Р4830/2, Р4830/3 — laboratory resistance boxes for DC and AC circuits. We reply within one business day.

* — required fields

Or email office@usk.ua or call +38 (044) 339-50-20.

REQUEST

Need a price, lead time or replacement?

Send us a request and we will confirm the configuration, lead time and delivery terms. We reply by email.

How we use your data

UKRSK, Ukrainian Components Warehouse, is a USK project. This form saves your email, request details, optional name and company, and attached files on the UKRSK server to process your enquiry. Files are not published in the catalogue. The form currently does not perform automated technical matching or send commercial quotations. To enquire about or delete your data, submit a request using this form and include your reference number. Do not attach passwords, payment details or unrelated confidential materials. Technical rate-limit records are stored separately; no advertising trackers are used.

To monitor page performance we collect LCP, INP and CLS measurements, language, screen category and a random measurement ID. These are stored on our hosting for up to 35 days; form contents, email, IP address and full URL are not recorded in this log. No cookies are used; Do Not Track and Global Privacy Control disable measurement.