Ukrainian Components WarehouseA USK project
UKRSKINDUSTRIAL COMPONENTS

TEST AND MEASUREMENT

Р5054 — conductance box for loading voltage transformers during verification

error +4 % · 50 or 60 Hz · U nom 100/√3, 100, 150 V · steps 1.25–25 VA

The Р5054 conductance box is a laboratory standard that applies a defined burden to the secondary circuit of a voltage transformer while it is verified on a differential-zero apparatus. The unit sets the required apparent power in switchable steps and provides the active and reactive components at a rated power factor of 0.8 or 1. The permissible basic error of the active and reactive components of power or conductance is +4 %. Versions are available for rated voltages of 100/√3 V, 100 V and 150 V. Р5054/1, /2 and /3 are for 50 Hz AC circuits, and Р5054/4, /5 and /6 are for 60 Hz. The 100/√3 V and 100 V versions offer steps from 1.25 to 25 VA. Depending on the version, the voltage may vary within 80–120 % or 20–120 % of nominal. Typical users are calibration laboratories, utility metrology departments and factory test bays that check voltage transformers.

Purpose
burden for voltage transformers during verification on a differential-zero apparatus
Permissible basic error (active and reactive components)
+4 %
Frequency
50 Hz (Р5054/1, /2, /3); 60 Hz (Р5054/4, /5, /6)
Rated power factor
0.8; 1

Technical specifications

Purposeburden for voltage transformers during verification on a differential-zero apparatus
Permissible basic error (active and reactive components)+4 %
Frequency50 Hz (Р5054/1, /2, /3); 60 Hz (Р5054/4, /5, /6)
Rated power factor0.8; 1
U nom, Р5054/1, /4100/√3 V
U nom, Р5054/2, /5100 V
U nom, Р5054/3, /6150 V
Voltage variation range, Р5054/1, /480–120 % of U nom
Voltage variation range, Р5054/2, /520–120 % of U nom
Apparent power steps at PF 0.8 (Р5054/1, /2, /4, /5)1.25; 1.67; 2.5; 5; 10; 15; 25; 25 VA
Apparent power steps at PF 1 (Р5054/1, /2, /4, /5)1.25; 2.5; 5; 10; 15 VA
Apparent power steps (Р5054/3, /6)1.25; 2.5; 5; 10; 15 VA
Overall dimensions, max490x130x380 mm (Р5054/1, /2, /4, /5); 490x130x260 mm (Р5054/3, /6)
Weight, max13 kg (Р5054/1, /2, /4, /5); 8 kg (Р5054/3, /6)

Specifications are taken from the manufacturer’s technical documentation. Updated .

Description

How to choose

Choose the version by the secondary voltage of the transformer under test and the supply frequency. For 100/√3 V, use Р5054/1 (50 Hz) or Р5054/4 (60 Hz), with a range of 80–120 % U nom. For 100 V, use Р5054/2 or Р5054/5, with a range of 20–120 % U nom. For 150 V, use Р5054/3 or Р5054/6. This lighter 8 kg version has steps of 1.25–15 VA.

Similar products

Test and measurement

Request a price and lead time

Р5054 — conductance box for loading voltage transformers during verification. We reply within one business day.

* — required fields

Or email office@usk.ua or call +38 (044) 339-50-20.

REQUEST

Need a price, lead time or replacement?

Send us a request and we will confirm the configuration, lead time and delivery terms. We reply by email.

How we use your data

UKRSK, Ukrainian Components Warehouse, is a USK project. This form saves your email, request details, optional name and company, and attached files on the UKRSK server to process your enquiry. Files are not published in the catalogue. The form currently does not perform automated technical matching or send commercial quotations. To enquire about or delete your data, submit a request using this form and include your reference number. Do not attach passwords, payment details or unrelated confidential materials. Technical rate-limit records are stored separately; no advertising trackers are used.

To monitor page performance we collect LCP, INP and CLS measurements, language, screen category and a random measurement ID. These are stored on our hosting for up to 35 days; form contents, email, IP address and full URL are not recorded in this log. No cookies are used; Do Not Track and Global Privacy Control disable measurement.