Ukrainian Components WarehouseA USK project
UKRSKINDUSTRIAL COMPONENTS

TEST AND MEASUREMENT

С196 — electrostatic kilovoltmeter for DC and AC voltage

0–7.5; 0–15; 0–30 kV · error ±1.0% · 20 Hz – 14 MHz · input capacitance ≤ 15 pF

The С196 is an electrostatic kilovoltmeter for measuring high DC and AC voltage across a wide frequency band. Its electrostatic measuring system places almost no load on the circuit under test: input capacitance does not exceed 15 pF, and the natural resonant frequency is at least 100 MHz. The instrument offers three indication ranges: 0–7.5, 0–15 and 0–30 kV. Measurements are taken between 27 and 100% of the range end value. The intrinsic error limit is ±1.0%, and settling time is no more than 4 s. The operating frequency band starts at 20 Hz and extends to 14, 8 or 2 MHz, depending on the range. Power is needed only for the scale illuminator, which runs from 127 or 220 V 50 Hz mains or from a 6 V source. The kilovoltmeter is used to measure voltage in circuits with one earthed pole. It is also suitable for unearthed circuits where both conductors are live relative to earth. The instrument is repairable and has an average service life of at least 12 years.

Indication ranges
0–7.5; 0–15; 0–30 kV
Measuring range
27–100% of the indication range end value
Current type
DC and AC
Reference frequency
50 Hz

Technical specifications

Indication ranges0–7.5; 0–15; 0–30 kV
Measuring range27–100% of the indication range end value
Current typeDC and AC
Reference frequency50 Hz
Operating frequency band20 Hz to 14 MHz (0–7.5 kV); to 8 MHz (0–15 kV); to 2 MHz (0–30 kV)
Intrinsic error limit±1.0% of range end value
Additional temperature error±1.0% per 10 °C (within 10–35 °C)
Input capacitancemax. 15 pF
Natural resonant frequencymin. 100 MHz
Settling timemax. 4 s
Insulation test voltage (electrode to case)16 kV (0–7.5 kV); 26 kV (0–15 kV); 45 kV (0–30 kV)
Illuminator circuit test voltage to case2 kV
High-voltage electrode insulation resistancemin. 104 MΩ
Illuminator circuit insulation resistancemin. 40 MΩ
Illuminator supply127 or 220 V AC 50 Hz; 6 V DC or AC 50 Hz
Average service lifemin. 12 years
Verification intervalonce a year
Dimensions280 × 645 × 239 mm
Weightmax. 11 kg
Scope of supplykilovoltmeter, power cord, ball attachment, spare lamp ОП6,3-0,22, technical description and operating manual, passport

Specifications are taken from the manufacturer’s technical documentation. Updated .

Description

How to choose

Select the range according to the expected voltage so that the reading falls within 27–100% of the range end value. For high-frequency measurements, observe the frequency limits: only the 0–7.5 kV range works up to 14 MHz, the 0–15 kV range up to 8 MHz, and the 0–30 kV range up to 2 MHz.

Similar products

Test and measurement

Request a price and lead time

С196 — electrostatic kilovoltmeter for DC and AC voltage. We reply within one business day.

* — required fields

Or email office@usk.ua or call +38 (044) 339-50-20.

REQUEST

Need a price, lead time or replacement?

Send us a request and we will confirm the configuration, lead time and delivery terms. We reply by email.

How we use your data

UKRSK, Ukrainian Components Warehouse, is a USK project. This form saves your email, request details, optional name and company, and attached files on the UKRSK server to process your enquiry. Files are not published in the catalogue. The form currently does not perform automated technical matching or send commercial quotations. To enquire about or delete your data, submit a request using this form and include your reference number. Do not attach passwords, payment details or unrelated confidential materials. Technical rate-limit records are stored separately; no advertising trackers are used.

To monitor page performance we collect LCP, INP and CLS measurements, language, screen category and a random measurement ID. These are stored on our hosting for up to 35 days; form contents, email, IP address and full URL are not recorded in this log. No cookies are used; Do Not Track and Global Privacy Control disable measurement.