С197 — electrostatic kilovoltmeter for DC and AC voltage
Country: Ukraine · Model: С197
0–7.5; 0–15; 0–30 kV · error ±1.0 % · 20 Hz – 10 MHz · input capacitance ≤ 15 pF
The С197 is an electrostatic kilovoltmeter for high DC and AC voltages. It operates in circuits with one earthed pole and in unearthed circuits where both conductors are live with respect to earth. Its electrostatic movement places almost no load on the circuit under test. Input capacitance is no more than 15 pF, and the natural resonant frequency is at least 100 MHz. The instrument has indication ranges of 0–7.5, 0–15 and 0–30 kV. Each range is usable from 27 to 100 % of its full-scale value. The limit of intrinsic error is ±1.0 % of full scale. The reference frequency is 50 Hz, and the AC operating band runs from 20 Hz up to 10 MHz. Pointer settling time does not exceed 4 s. A scale illuminator is powered from 127 or 220 V or from a 6 V supply. Typical uses are high-voltage laboratories, test benches, and insulation and HV equipment testing.
- Indication ranges
- 0–7.5; 0–15; 0–30 kV
- Measuring range
- 27–100 % of full-scale value
- Current type
- DC and AC
- Circuit type
- earthed and unearthed
Technical specifications
| Indication ranges | 0–7.5; 0–15; 0–30 kV |
|---|---|
| Measuring range | 27–100 % of full-scale value |
| Current type | DC and AC |
| Circuit type | earthed and unearthed |
| Reference frequency | 50 Hz |
| Operating frequency band | 20 Hz – 10 MHz (0–7.5 kV); 20 Hz – 8 MHz (0–15 kV); 20 Hz – 5 MHz (0–30 kV) |
| Extended frequency band | above 10 up to 14 MHz (0–7.5 kV range) |
| Intrinsic error limit | ±1.0 % of full scale |
| Additional temperature error | ±0.5 % per 10 °C (within 10–35 °C) |
| Input capacitance | ≤ 15 pF |
| Natural resonant frequency | ≥ 100 MHz |
| Settling time | ≤ 4 s |
| Insulation test voltage (electrode to case) | 16 kV (0–7.5 kV); 26 kV (0–15 kV); 45 kV (0–30 kV) |
| Test voltage (illuminator circuit to case) | 2 kV |
| Illuminator circuit insulation resistance | ≥ 40 MΩ |
| Illuminator supply | 50 Hz AC 127 or 220 V; 6 V DC or 50 Hz AC |
| Mean service life | at least 12 years |
| Overall dimensions | 250 × 490 × 230 mm |
| Weight | ≤ 8.5 kg |
| Scope of supply | instrument, power cord, spare lamp ОП6,3-0,22, technical description, passport |
Specifications are taken from the manufacturer’s technical documentation. Updated .
Description
How to choose
Choose a range in which the measured voltage falls between 27 and 100 % of full scale. The 0–7.5 kV range reaches 10 MHz, with an extended band up to 14 MHz. The 0–15 kV range reaches 8 MHz and the 0–30 kV range 5 MHz. The illuminator runs from 127 or 220 V mains or from a 6 V DC or AC supply.
Similar products
Request a price and lead time
С197 — electrostatic kilovoltmeter for DC and AC voltage. We reply within one business day.
Or email office@usk.ua or call +38 (044) 339-50-20.
REQUEST
Need a price, lead time or replacement?
Send us a request and we will confirm the configuration, lead time and delivery terms. We reply by email.