Sharky FS 473 — ultrasonic flow transducers for heat meters DN15–100, qp 0.6–60 m³/h, up to 150 °C

Sharky FS 473 is an ultrasonic flow transducer for compact heat and cold meters in DN15–100 with nominal flows qp 0.6–60 m³/h. It works with a calculator, for example FAUN, and measures water from 5 °C to 130 °C (qp 0.6–2.5) or up to 150 °C (qp 3.5–60) at pressure up to 16 bar. It suits heating substations, metering stations of apartment and public buildings, and industrial heat consumers.

Type
ultrasonic flow transducer for compact heat/cold meters
Nominal flow qp
0.6; 1.5; 2.5; 3.5; 6; 10; 15; 25; 40; 60 m³/h
Sizes
DN15–100 (thread G¾B–G2B, flanges DN32–100)
Water temperature
5–130 °C (qp 0.6–2.5); 5–150 °C (qp 3.5–60)

Technical specifications

Typeultrasonic flow transducer for compact heat/cold meters
Nominal flow qp0.6; 1.5; 2.5; 3.5; 6; 10; 15; 25; 40; 60 m³/h
SizesDN15–100 (thread G¾B–G2B, flanges DN32–100)
Water temperature5–130 °C (qp 0.6–2.5); 5–150 °C (qp 3.5–60)
PressurePN16
Dynamic rangeqi/qp 1:100; 1:250 (class 2) for qp 1.5–25
Accuracy classEN 1434 / MID class 2
Environment classEN 1434 C, MID E2 + M2; ambient 5–55 °C
ProtectionIP54 (heating), IP65 (cooling)
Supplyfrom the calculator (3.0 V battery, up to 12 years) or external 3.0–5.5 V DC
Outputpulse (open collector), 1 ml … 5 000 l/pulse, cable 2.4 m (optional 4.9/9.9 m)
Length / weight110–360 mm / 0.6–16.9 kg; body brass (qp 0.6–10) or cast iron (qp 15–60)

Specifications are taken from the manufacturer’s technical documentation. Updated .

Description

The ultrasonic method gives stable measurements, low pressure loss, a low start-up threshold and a small minimum flow, and the instrument is insensitive to dirty water and scale. There are no moving parts, installation is permitted in any position, and straight sections before and after the transducer are not needed. The standard dynamic range is qi/qp 1:100, and for qp 1.5, 2.5, 6, 10, 15 and 25 m³/h a 1:250 range in class 2 is available on request. The range covers qp 0.6 (DN15, G¾B, 110 mm), 1.5 (DN15/20), 2.5 (DN20, G1B, 130 or 190 mm), 3.5 and 6 (DN25, G1¼B, 260 mm, and for 6 also a flanged DN32), 10 (DN40, G2B 200 mm or flange 300 mm), 15 (DN50), 25 (DN65), 40 (DN80) and 60 m³/h (DN100, 360 mm, 16.9 kg). The output is a pulse output of open-collector type with pulse weights of 1, 10 and 100 l/pulse, from 1 ml to 5 000 l/pulse depending on DN, cable 2.4 m (optionally 4.9 or 9.9 m). The environment class is EN 1434 C, MID E2 + M2, protection IP54 for heating and IP65 for cooling.

How to choose

The model is chosen by nominal flow qp and size. The 1:250 range in class 2 on request is available only for qp 1.5–25. If the water is hotter than 130 °C, the qp 3.5–60 versions are needed, which work up to 150 °C.

Request a price and lead time

Sharky FS 473 — ultrasonic flow transducers for heat meters DN15–100, qp 0.6–60 m³/h, up to 150 °C. We reply within one business day.

* — required fields

Or email office@usk.ua or call +38 (044) 339-50-20.

FAQ

Up to what temperature does Sharky FS 473 work?

From 5 °C to 130 °C for qp 0.6–2.5 and up to 150 °C for qp 3.5–60. Pressure up to 16 bar.

Are straight sections needed before and after Sharky FS 473?

No, straight sections are not needed, and installation is permitted in any position. A magnetic field does not affect operation.

What is the dynamic range of Sharky FS 473?

Standard qi/qp 1:100, and for qp 1.5–25 on request 1:250 in class 2.

What supply and output does the transducer have?

Supply from the calculator (3.0 V battery, up to 12 years) or from an external 3.0–5.5 V DC source. The output is a pulse output of open-collector type.

Similar products

Test and measurement

REQUEST

Need a price, lead time or replacement?

Send us a request and we will confirm the configuration, lead time and delivery terms. We reply by email.

How we use your data

UKRSK, Ukrainian Components Warehouse, is a USK project. This form saves your email, request details, optional name and company, and attached files on the UKRSK server to process your enquiry. Files are not published in the catalogue. The form currently does not perform automated technical matching or send commercial quotations. To enquire about or delete your data, submit a request using this form and include your reference number. Do not attach passwords, payment details or unrelated confidential materials. Technical rate-limit records are stored separately; no advertising trackers are used.

To monitor page performance we collect LCP, INP and CLS measurements, language, screen category and a random measurement ID. These are stored on our hosting for up to 35 days; form contents, email, IP address and full URL are not recorded in this log. No cookies are used; Do Not Track and Global Privacy Control disable measurement.